The SciPres combines pressure sensing capabilities and the convenience of disposability with easy setup. Each sensor is preprogrammed and barcoded with a unique ID for easy traceability and data documentation when combined with the SciLog SciDoc software. Factory calibration data is also stored on each sensor's chip for out-of-box, plug and play use. The SciPres comes in 5 different sizes to fit a variety of tubing sizes: Luer, 3/8" barb, 1/2" barb, 3/4" TC (Tri-Clover), and 1.0" TC (Tri-Clover).
SciPres Pressure Monitor
The SciPres Sensor Monitor provides three sensor inputs for measuring up to three SciLog pressure sensors (P1, P2 & P3) simultaneously. The sensor readout permits the user selectable display of differential pressure (DP) and trans-membrane pressure (TMP). The SciPres Monitor has both analog outputs and digital outputs (RS-232). Four TTL alarm switches and user-selectable high / low pressure limit settings for each sensor input.
![]()
SciPres Product Information
- SciPres Part Numbers & Specs
-
SciPres Sensors
Connector Part Number Compatible Tubing Sizes Max Flow Rate Max PSI Luer 080-699PSX-5 ID 0.03" to 0.31" 1 liters/min 60 psi 3/8" Barb 080-694PSX-5 ID 0.31" to 0.38" 8 liters/min 60 psi 1/2" Barb 080-695PSX-5 ID 0.50" 17 liters/min 60 psi 3/4" Tri-Clover (TC) 080-696PSX-5 Tubing with 3/4" TC 31 liters/min 60 psi 1.0" TC "Ladish" 080-697PSX-5 Tubing with 1" TC Ladish 60 liters/min 60 psi SciPres Monitors
Voltage Part Number Form Factor 110 VAC 080-690 Bench-Top Monitor 220 VAC 081-690 Bench-Top Monitor 24 VDC 080-691 DIN Rail Mount - SciPres Sensing Element Specs
-
- All wetted materials are made of animal-free compounds.
- Material, Fluid Contact: Natural Polysulfone meets USP Class VI and FDA 21 CFR 177.1520.
- Compatible with most sanitizing agents such as NaOH for flow cell sanitization.
- Sensor Type: Medical Grade, Piezoresistive Sensing Element with on-chip temperature compensation
- Sensor Isolation: Insoluble silicone dielectric gel isolates sensing element from process solution. The gel is a non-toxic, non-allergenic elastomer system that meets all USP XX Biological Class V requirements.
- Pressure Range: 0 - 60 psi
- Pressure Accuracy: ± 0.30 psi at 60 psi
- Pressure Resolution: 0.01 psi
- Temperature Range: 0-60° C
- Sensor Microchip: EPROM, Stored Sensor ID and Calibration Factor
- Sensor Connector Cables: Lockable & Waterproof
- Sanitization: Can be sanitized with NaOH, NaCl, Formaldehyde, etc. Also Autoclaved
- SciPres Monitor Specs
-
- Sensor Inputs: Up to Three SciLog Pressure Sensors (P1, P2, P3) Simultaneously.
- Sensor Readout: Display of Sensor P1, P2 and P3 as well as Differential Pressure: DP= (P1 - P2) and Transmembrane
- Pressure: TM= [(P1+P2/2)-P3]. Display of DP and TM are User-Selectable.
- Analog Outputs: 4-20 mA Outputs for P1, P2, P3, and DP or TM, 18 Bit Resolution.
- Digital Outputs: RS-232 and RS-485 for networking.
- Alarm Outputs: 4 TTL Switches, User-Selectable Hi/Lo Pressure Limit Settings for P1, P2, P3, and DP or TM.
- Power Supply: Wall Transformer (Lab Option); 115/230 VAC (Process Option).
- Display: 2 Lines, 20 Characters each, Backlit.
- Data Entry: Membrane Switch, 6 Keys.
SOFTWARE - SELECTABLE PARAMETERS
- Pressure Units: psi, kPa, bar, mmHg.
- Pressure Range, High: 100 psi, Analog Out at 100%.
- Pressure Range, Low: 0.0 psi, Analog Out at 0%.
- Serial Com: BR=9600, WL=8, SB=2, PT=N.
- Print Time: User Selectable, 1 to 60 Seconds.
- Display Mode: Transmembrane Pressure (TM) or Differential Pressure (DP) as well as Display of P1, P2 and P3.
- Alarm Outputs:
- P1 Hi/Lo
- P2 Hi/Lo
- P3 Hi/Lo
- TM/DP Hi/Lo
- Time/Date: All Data Printout with Time/Date Stamp.
- Lock-Out: On/Off, Requires Password (optional).
- Re-Cal: Custom Calibration of Sensors.
- Material Certificates
-
- BSE Material Compliance Statement SciLog Disposable Sensors Rev A AD080610.pdf
- Material Compliance Statement Concerning Use of Latex SciLog Disposable Sensors.pdf
- USP Biological Reactivity Tests In Vivo Material Compliance Statement SciLog Disposable Sensors.pdf
- USP Class VI Material Compliance Statement SciLog Disposable Sensors Rev A AD080610.pdf
- Performance Data
-
- Out-of-Box Performance SciPres 1.0 Inch.pdf
- Out-of-Box Performance SciPres Luer.pdf
- SciPres 100 Hour Exposure pH 2 Sulfuric Acid.pdf
- SciPres 100 Hour Exposure to 1.0M NaOH.pdf
- SciPres 100 Hour Exposure to 70 Per Cent IPA.pdf
- SciPres FS Accuracy and Precision.pdf
- SciPres Post-Autoclave Response.pdf
- SciPres Post-Gamma Accuracy.pdf
- SciPres Sanitization 1.0N NaOH at 22C and 50C.pdf
- SciPres Sensor Sanitization Summary.pdf
- Regulatory Compliance